Newer
Older
##### _DStat is described in detail in [Dryden MDM, Wheeler AR (2015) DStat: A Versatile, Open-Source Potentiostat for Electroanalysis and Integration. PLoS ONE 10(10): e0140349. doi: 10.1371/journal.pone.0140349](http://journals.plos.org/plosone/article?id=10.1371/journal.pone.0140349) If you use this information in published work, please cite accordingly._
---
This is the documentation for the DStat interface software.
The DStat interface is written primarily in Python and runs on Linux, Mac, and Windows.
It is the main method for running experiments on the DStat, controlling experimental parameters and collecting and plotting data.
It currently has no abilities for analyzing recorded data or opening previously saved data files, but data is saved in a simple text format or numpy-compatible binary format and plots can be saved as images.
## Table of Contents:
1. [Installation](#Installation)
1. [MacOS](#macos)
1. [Using Anaconda (Recommended)](#using-anaconda-recommended)
2. [Old Homebrew Instructions](#old-homebrew-instructions)
1. [Linux](#linux)
2. [Windows](#windows)
2. [Getting Started](#Getting-Started)
# Introduction
The DStat interface is written primarily in Python and runs on Linux, Mac, and Windows.
It is the main method for running experiments on the DStat, controlling experimental parameters and collecting and plotting data.
It currently has no abilities for analyzing recorded data or opening previously saved data files, but data is saved in a simple text format and plots can be saved as images.
dstat-interface has moved to gtk+3 and we now recommend Anaconda/Miniconda for installation.
## MacOS
### Using Anaconda (Recommended)
1. [Install Miniconda](https://repo.continuum.io/miniconda/Miniconda2-latest-MacOSX-x86_64.sh) It doesn't matter if you pick Python 2.7 or 3.5—this just sets Miniconda's default Python. (Skip if Miniconda or Anaconda are already installed)
2. Create a new environment for dstat. In a terminal type:
````conda create -n dstat -c mdryden python=2.7 dstat-interface dstat-interface-deps````
3. Then to run dstat-interface:
````source activate dstat
python -m dstat_interface.main````
The easiest way to get most of the necessary requirements to run dstat-interface is using [Homebrew](http://brew.sh):
brew tap homebrew/python
brew update
brew install python gobject-introspection gtk+3 pygobject3 py2cairo scipy zeromq
brew install matplotlib --with-pygtk
Be patient on the last step—matplotlib needs to be compiled and may take 2 or 3 minutes.
Make sure you're using brew-installed python, not OS X's default python. `which python` should point to `/usr/local/bin/python` not `/usr/bin/python`. Type `brew doctor` for more information if you are having issues.
The final requirements, can be installed using python's pip system:
pip install pandas pyserial pyzmq pyyaml seaborn zmq-plugin
## Linux
Linux prerequisite installation is similar to that of MacOS with Homebrew, only using your distribution's native package manager rather than Homebrew, and X11 will likely be installed already. Some distributions may not have packages available for installing matplotlib or numpy, in which case, they should be installed using pip.
The final requirements, can be installed using python's pip system:
pip install pandas pyserial pyzmq pyyaml seaborn zmq-plugin
## Windows
The following terminal commands will result in a full installation of dstat-interface and its requirements, assuming [64-bit Miniconda][1] is installed:
````
conda create -n dstat -c mdryden python=2 dstat-interface
activate dstat
````
To finish the installation, GTK+3 and its Python bindings must be installed:
1. Download the latest all-in-one installer from [here][2].
2. When the installer prompts for the path to your Python distribution, navigate to your Miniconda/Anaconda envs folder and choose the dstat folder. If you chose to install Miniconda for your user only, the envs folder is found in `$USER/Miniconda2/envs` (for Miniconda) or `$USER/Anaconda/envs` (for Anaconda), where `$USER` is your user directory. If you installed Miniconda for all users, the Miniconda2/Anaconda folder will be in the root of your C: drive.
3. When the installer asks which modules to install, choose GTK.
4. Finish the installer.
We are installing in a separate environment to keep a clean system.
`activate dstat` will enter the environment (must be done whenever a new terminal is opened),
and `deactivate` will return to the root environment.
Therefore, to run dstat-interface from our environment, we must first activate it (if not already done) before launching it:
activate dstat
python -m dstat_interface.main
[1]: https://repo.continuum.io/miniconda/Miniconda2-latest-Windows-x86_64.exe
[2]: https://sourceforge.net/projects/pygobjectwin32/
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
# Getting started
## Interface overview
![Main interface](images/1.png)
1. Menu bar
![Menu bar](images/5.png)
* File
* Save Current data… — Saves the data of the currently visible plot as a space-separated text file or numpy .npy file
* Save Plot… – Save the currently visible plot as a .pdf
* Quit — Quits dstat-interface
* Dropbot
* Connect — Listens for µDrop connection over ZMQ
* Disconnect — Disconnect from µDrop
* Help
* About — Displays license information
2. ADC Settings Panel
* PGA Setting — Sets the ADC's internal voltage gain. Should almost always be left at 2x except for potentiometry as increasing voltage gain reduces S/N. Settings other than 2x do not adjust data to match (e.g. one must halve measured current/voltage values if PGA is set to 4x).
* Sample Rate - Sets the ADC's sample frequency. Lower rates give less noise, but reduced temporal resolution. Digital filter has a zero at multiples of the sample rate, so the Sample Rate setting can be used to filter out AC line noise by setting the rate to a factor of the line frequency, e.g. for 60 Hz rejection, choose 60, 30, 15, 10, 5, or 2.5 Hz.
* Input Buffer — Sets the ADC's internal input buffer. Should generally be enabled.
3. Potentiostat Settings Panel
* Gain — Controls the current-to-voltage converter gain. Higher values produce better S/N but reduce full scale current limit. Has no effect on potentiometry experiments.
4. Experiment Panel - Pulldown menu changes between experiment types and parameters are entered below.
5. Experiment Control
* Execute — Start the currently selected experiment with the given parameters.
* Stop — Stop the currently running experiment. If Autosave is enabled, the partial experiment will be saved.
6. Communications Panel
* Serial Port — Select the port where DStat is located. On Windows, this is generally something like `COM3`. On Mac OS X, it should appear as `/dev/cu.usbmodem12...E1`. On Linux, it may vary, but will start with `/dev`. If you're not sure, the simplest way is to check the list before and after plugging the DStat in. (Clicking the Refresh button after)
* Refresh — Refreshes the Serial Port list
* Connect — Attempts to handshake with DStat. If unsuccessful, it will time out after approximately 30 seconds.
* OCP — Displays the current open circuit potential measured at the reference electrode input. Active when DStat is connected and an experiment is not running.
* Status bar — Displays status and error messages.
7. Data display tabs — Switches between the plot and raw data tabs
* Plot — Displays the graphical representation of the incoming data.
* Raw Data — The raw experiment data. Doesn't appear until the experiment is complete. The first column corresponds to the x-axis of the plot (time or voltage), and the second column corresponds to the y-axis (current or voltage). For mult-scan experiments, additional pairs of columns represent successive scans.
![raw data](images/4.png)
* Extra Data — For SWV and DPV, the separate forward and reverse currents are recorded here.
8. Autosave controls
* Autosave — Enables automatic data saving on experiment completion. A text data file and a .pdf image of the plot will be saved.
* File save location
* File name selector — A number will be appended automatically if a file with the same name already exists.
9. Plot display
10. Plot navigation controls — For changing the view of the data plot.
## Connecting to DStat
1. Plug the DStat into a USB port, ensuring that drivers are loaded correctly on Windows systems.
2. Click Refresh in the Communications Panel to refresh the Serial Port list and choose the correct entry for the DStat (described above).
3. Click Connect.
4. If the connection was successful, a number should appear in the OCP field and the version number will appear in the status bar.
![connect](images/2.png)
If the connection failed, unplug the DStat and try again.
## Running an experiment
1. Choose the experiment you want to run in the Experiment Panel.
2. Fill the parameter fields.
3. Set an appropriate potentiostat gain.
4. Click Execute.